- GC-TQMS Application: Simultaneous determination of residual agricultural chemicals in food by GC-MS/MS - Sensitivity of standard solution 1ppb and linearity of calibration curve-
GC-TQMS Application: Simultaneous determination of residual agricultural chemicals in food by GC-MS/MS - Sensitivity of standard solution 1ppb and linearity of calibration curve-
- Summary
MSTips No.259
As "food safety" is recognized as an increasingly important issue on a global scale, many nations have their own regulations on residual agricultural chemicals in food. In Japan, the positive list system, which was enforced at the end of May 2006, stipulates a uniform standard of 10 ppb as a quantity that is considered safe for human health. Under the positive list system, more agricultural chemicals need to be examined, and as a result, techniques capable of accurately and collectively analyzing residual agricultural chemicals in food are in increasing demand. While mass spectrometry (MS) is known for its high detection sensitivity, MS/MS is becoming the mainstream of pesticide analysis for its superior sensitivity and selectivity.
The JMS-TQ4000GC, JEOL’s latest GC-MS/MS system, has a unique ion storage/ejection mechanism within the MS/MS collision cell and incorporates new firmware to support MS/MS analysis with up to 36,000 transitions. In this work, we report the verified results for pesticide residues sensitivity in food using the JMS-TQ4000GC.
Sample and Method
A pesticide standard solution from FUJIFILM Wako Pure Chemical Corporation (PL series) was used that consisted of equal amounts of PL 1, 2, 3, 4, 5, 6, 11, and 12. Afterwards, the solution was diluted to 1, 5, 10, 50 and 100 ppb. PEG 300 was used to protect the pesticides from thermal decomposition in the GC injection liner. In this work, both the sensitivity and the linearity of the calibration curve were examined for 150 pesticides. Table 1 shows the measurement conditions used for the analysis.
Table 1 Measurement condition
[GC-TQMS condition]
System |
JMS-TQ4000GC (JEOL) |
Ionization mode | EI+: 70eV, 50μA |
GC column | VF-5ms, 30m x 0.25mm, 0.25μm |
Oven temp. | 50°C (1min)→25°C/min→125°C→10°C/min→300°C |
Inlet temp. | 250°C |
Inlet mode | Splitless, 2μL |
He flow | 1.0mL/min (Constant Flow) |
MS/MS mode | Peak Dependent SRM |
Results and Discussion
Figure 1 shows the calibration curves from 1-100ppb as well as the SRM chromatograms for 4 pesticides at 1ppb. The JMS-TQ4000GC showed excellent linearity and sensitivity for this range of concentrations. Fig 2 shows the retention times (R.T.), SRM information, and correlation coefficient for 150 pesticide calibration curves. Since the uniform criterion of the positive list system is 10 ppb, when the concentration ratio in the sample pretreatment is equal, the criteria concentration in the measurement sample for GC-MS/MS is also 10 ppb. A number of pesticides were detected with good sensitivity at 1 ppb which is 1/10 of criteria concentration. These results suggest that the JMS-TQ4000GC is the most effective tool for pesticide analysis.

Fig.1 Calibration curves (1-100ppb) and SRM chromatograms for 1ppb data
Table 2 Calibration curve information for 150 pesticides using 1, 5, 10, 50 and 100 ppb standard solutions (n=3 each)
No. | Compound | RT (min) | Q ion | Correlation coefficient, R (1-100ppb) | |
---|---|---|---|---|---|
Precursor ion | Product ion | ||||
1 |
Atrazine |
12.37 | 215 | 200 | 0.9999 |
2 | Benthiocarb | 14.68 | 100 | 72 | 0.9997 |
3 | cis-Permethrin | 20.68 | 183 | 153 | 0.9999 |
4 | Diflufenican | 18.31 | 266 | 218 | 0.9999 |
5 | Fenamiphos | 16.24 | 303 | 195 | 0.9998 |
6 | Fenarimol | 20.20 | 139 | 111 | 0.9999 |
7 | Fenpropimorph | 14.74 | 128 | 110 | 0.9998 |
8 | Norflurazon | 17.94 | 303 | 145 | 0.9997 |
9 | Oxadiazon | 16.50 | 175 | 112 | 0.9996 |
10 | Penconazole | 15.45 | 248 | 192 | 0.9999 |
11 | Pendimethalin | 15.29 | 252 | 162 | 0.9990 |
12 | Procymidone | 15.70 | 283 | 255 | 0.9980 |
13 | Spiroxamine 1 | 13.78 | 100 | 72 | 0.9997 |
14 | Spiroxamine 2 | 14.31 | 100 | 72 | 0.9995 |
15 | Tefluthrine | 12.92 | 177 | 127 | 0.9999 |
16 | Terbufos | 12.63 | 231 | 175 | 0.9998 |
17 | Terbutryn | 14.31 | 241 | 170 | 0.9999 |
18 | trans-Permethrin | 20.68 | 183 | 153 | 0.9999 |
19 | Alachlor | 13.84 | 188 | 160 | 0.9994 |
20 | Buprofezin | 16.68 | 172 | 115 | 0.9993 |
21 | cis-Chlorfenvinphos | 15.49 | 323 | 267 | 0.9998 |
22 | Cyproconazole 1 | 17.03 | 139 | 111 | 0.9998 |
23 | Cyproconazole 2 | 17.03 | 139 | 111 | 0.9998 |
24 | Difenoconazole 1 | 23.10 | 323 | 265 | 0.9998 |
25 | Difenoconazole 2 | 23.18 | 323 | 265 | 0.9999 |
26 | Ethion | 17.34 | 231 | 175 | 0.9998 |
27 | Fenitrothion | 14.33 | 277 | 260 | 0.9989 |
28 | Fenthion | 14.73 | 278 | 245 | 0.9997 |
29 | Fluridone | 22.18 | 328 | 259 | 0.9997 |
30 | Hexazinone | 18.24 | 171 | 71 | 0.9997 |
31 | Isofenphos oxon | 14.79 | 229 | 201 | 0.9999 |
32 | Isophenphos | 15.43 | 255 | 185 | 0.9998 |
33 | Isoprothiolane | 16.44 | 189 | 145 | 0.9997 |
34 | Propargite 1 | 18.31 | 135 | 107 | 0.9996 |
35 | Propargite 2 | 18.31 | 135 | 107 | 0.9996 |
36 | Propiconazole 1 | 17.93 | 173 | 145 | 0.9998 |
37 | Propiconazole 2 | 18.07 | 173 | 145 | 0.9996 |
38 | Propyzamide | 12.73 | 175 | 147 | 0.9998 |
39 | Pyriproxyfen | 19.72 | 136 | 96 | 0.9999 |
40 | trans-Chlorfenvinphos | 15.25 | 323 | 267 | 0.9993 |
41 | Triadimenol 1 | 15.70 | 168 | 70 | 0.9997 |
42 | Triadimenol 2 | 15.84 | 168 | 70 | 0.9994 |
43 | Triallate | 13.13 | 268 | 184 | 0.9999 |
44 | Vinclozoline | 13.76 | 212 | 172 | 0.9997 |
45 | Acetamiprid | 18.92 | 152 | 116 | 0.9997 |
46 | Allethrin 1 | 15.43 | 123 | 81 | 0.9994 |
47 | Allethrin 2 | 15.51 | 123 | 81 | 0.9991 |
48 | Bitertanol 1 | 20.66 | 170 | 141 | 0.9999 |
49 | Bitertanol 2 | 20.76 | 170 | 141 | 0.9997 |
50 | Bromopropylate | 19.00 | 183 | 155 | 0.9998 |
51 | Chlorobenzilate | 17.19 | 251 | 139 | 0.9998 |
52 | Chlorpyrifos | 14.63 | 197 | 169 | 0.9997 |
53 | Oxyfluorfen | 16.59 | 252 | 196 | 0.9955 |
54 | Parathion | 14.80 | 291 | 109 | 0.9979 |
55 | Pirimiphos methyl | 14.24 | 290 | 233 | 0.9998 |
56 | Propanil | 13.67 | 217 | 161 | 0.9999 |
57 | Pyridaben | 20.87 | 147 | 119 | 0.9999 |
58 | Quinoxyfen | 18.00 | 237 | 208 | 0.9998 |
59 | Simazine | 12.29 | 201 | 173 | 0.9999 |
60 | Tebuconazole | 18.38 | 250 | 125 | 0.9998 |
61 | Triadimefon | 14.87 | 208 | 181 | 0.9993 |
62 | Triazophos | 17.65 | 161 | 134 | 0.9998 |
63 | Ametryn | 14.00 | 227 | 170 | 0.9999 |
64 | Azaconazole | 16.83 | 217 | 173 | 0.9998 |
65 | Bupirimate | 16.66 | 273 | 193 | 0.9998 |
66 | Butachlor | 16.01 | 176 | 147 | 0.9994 |
67 | Chlorthal dimethyl | 14.73 | 299 | 221 | 0.9999 |
68 | Dicloran | 12.19 | 206 | 176 | 0.9993 |
69 | Diethofencarb | 14.63 | 267 | 225 | 0.9999 |
70 | Dimepiperate | 15.74 | 119 | 91 | 0.9997 |
71 | Dimethenamid | 13.61 | 230 | 154 | 0.9999 |
72 | Etoxazole | 19.03 | 204 | 176 | 0.9998 |
73 | Fluacrypyrim | 17.38 | 204 | 189 | 0.9999 |
74 | Lenacil | 18.14 | 153 | 136 | 0.9998 |
75 | Pyriminobac methyl 1 | 17.96 | 302 | 256 | 0.9996 |
76 | Pyriminobac methyl 2 | 17.13 | 302 | 256 | 0.9997 |
77 | Pyroquilon | 12.88 | 173 | 130 | 0.9996 |
78 | Tetradifon | 19.53 | 159 | 131 | 0.9999 |
79 | Tolclofos-methyl | 13.87 | 265 | 250 | 0.9998 |
80 | Uniconazole P | 16.61 | 234 | 165 | 0.9996 |
81 | Acetochlor | 13.66 | 146 | 131 | 0.9997 |
82 | Benalaxyl | 17.83 | 204 | 176 | 0.9996 |
83 | Benfuresate | 13.57 | 256 | 163 | 0.9998 |
84 | Cadusafos | 11.71 | 159 | 131 | 0.9997 |
85 | Chlorpropham | 11.37 | 213 | 171 | 0.9997 |
86 | Diclocymet 1 | 15.61 | 277 | 221 | 0.9995 |
87 | Diclocymet 2 | 15.91 | 277 | 221 | 0.9999 |
88 | Dimethametryn | 15.40 | 212 | 122 | 0.9997 |
89 | Esprocarb | 14.49 | 222 | 91 | 0.9998 |
90 | Etofenprox | 21.83 | 163 | 135 | 0.9998 |
91 | Fenothiocarb | 16.08 | 160 | 72 | 0.9996 |
92 | Iprobenfos | 13.27 | 204 | 91 | 0.9998 |
93 | Isoxathion | 16.92 | 177 | 130 | 0.9997 |
94 | Mepronil | 17.61 | 119 | 91 | 0.9998 |
95 | Prometryn | 14.04 | 241 | 184 | 0.9999 |
96 | Propachlor | 10.85 | 120 | 92 | 0.9996 |
97 | Prothiofos | 16.41 | 267 | 239 | 0.9990 |
98 | Pyrifenox 1 | 15.99 | 262 | 200 | 0.9998 |
99 | Pyrimethanil | 12.89 | 198 | 183 | 0.9998 |
100 | Simetryn | 13.94 | 213 | 170 | 0.9999 |
101 | Terbacil | 13.06 | 161 | 144 | 0.9999 |
102 | Tetraconazole | 14.82 | 336 | 218 | 0.9998 |
103 | Thenylchlor | 18.31 | 127 | 99 | 0.9994 |
104 | Tribufos | 16.58 | 202 | 147 | 0.9995 |
105 | Tricyclazole | 16.66 | 189 | 161 | 0.9996 |
106 | Zoxamide (decomposed) | 15.72 | 187 | 159 | 0.9992 |
107 | Benoxacor | 13.40 | 259 | 120 | 0.9995 |
108 | Bromacil | 14.44 | 205 | 188 | 0.9996 |
109 | Bromobutide | 13.70 | 119 | 91 | 0.9998 |
110 | Butamifos | 16.17 | 286 | 202 | 0.9997 |
111 | Dichlofenthion | 13.56 | 279 | 223 | 0.9998 |
112 | Diphenamid | 15.12 | 167 | 152 | 0.9999 |
113 | Hexaconazole | 16.42 | 214 | 172 | 0.9996 |
114 | Mefenoxam | 14.00 | 206 | 132 | 0.9998 |
115 | Napropamide | 16.33 | 128 | 72 | 0.9997 |
116 | Oxadixyl | 17.40 | 163 | 132 | 0.9998 |
117 | Paclobutrazol | 16.07 | 236 | 125 | 0.9998 |
118 | Phenothrin 1 | 19.31 | 183 | 153 | 0.9998 |
119 | Phenothrin 2 | 19.42 | 183 | 153 | 0.9999 |
120 | Piperophos | 18.98 | 320 | 122 | 0.9998 |
121 | Prohydrojasmon 1 | 12.94 | 153 | 97 | 0.9998 |
122 | Prohydrojasmon 2 | 13.23 | 153 | 97 | 0.9996 |
123 | Propazine | 12.42 | 214 | 172 | 0.9997 |
124 | Pyributicarb | 18.63 | 165 | 108 | 0.9998 |
125 | Pyrifenox 2 | 15.51 | 262 | 200 | 0.9995 |
126 | Quinalphos | 15.63 | 146 | 118 | 0.9999 |
127 | Tebufenpyrad | 19.16 | 333 | 171 | 0.9998 |
128 | Tolfenpyrad | 24.05 | 197 | 154 | 0.9999 |
129 | Aldrin | 14.78 | 263 | 193 | 0.9988 |
130 | cis-Chlordane | 16.26 | 375 | 266 | 0.9998 |
131 | Dicofol | 4.97 | 139 | 111 | 0.9998 |
132 | Dieldrin | 16.78 | 277 | 206 | 0.9988 |
133 | Endrin | 17.19 | 263 | 193 | 0.9996 |
134 | Heptachlor | 14.07 | 272 | 237 | 0.9997 |
135 | trans-Chlordane | 16.02 | 373 | 266 | 0.9998 |
136 | 1-Naphthylacetamide | 14.40 | 141 | 115 | 0.9999 |
137 | Bromophos ethyl | 15.87 | 357 | 301 | 0.9998 |
138 | Carboxin | 16.80 | 235 | 143 | 0.9999 |
139 | Chlorbenside | 15.98 | 268 | 125 | 0.9995 |
140 | Chlorofenson | 16.42 | 175 | 111 | 0.9995 |
141 | clomazone | 12.47 | 125 | 89 | 0.9998 |
142 | Disulfoton | 12.98 | 88 | 60 | 0.9997 |
143 | Epoxiconazole | 18.61 | 192 | 138 | 0.9999 |
144 | Ethofumesate | 14.37 | 207 | 161 | 0.9997 |
145 | Fenamidone | 19.17 | 268 | 180 | 0.9998 |
146 | Flutriafol | 16.28 | 219 | 123 | 0.9999 |
147 | Isazophos | 12.96 | 161 | 119 | 0.9998 |
148 | Phorate | 11.79 | 121 | 65 | 0.9998 |
149 | Picolinafen | 18.97 | 376 | 238 | 0.9995 |
150 | Propaphos | 15.89 | 220 | 140 | 0.9998 |
- If you want to see a printed version, please click this PDF file.