NMR software DeltaV5.3.1 Introduction of new features ~ "pure shift" (1) ~PSYCHE

  • Summary

NM180008

"pure shift" is a method for homo nuclear broad band decoupling. Sensitivity would be reduced, but spectra of only (pure) chemical shifts can be obtained by eliminating j coupling.
The JNM - ECZ series has been adapted for "pure shift" measurement & processing from NMR software Delta V 5.3.1.
The following is an example of PSYCHE measurement for "pure shift".

pure shift JNM-ECZ series

Instrument example: JNM-ECZ400S and ROYALPROBE™

※Please note that the design, specifications, appearance, price of the product may be changed without notice.
Mathias Nilsson, and Gareth A. Morris, Angew. Chem. Int. Ed. 2014, 53, 1–4

Please see the PDF file for the additional information.
Another window opens when you click.

PDF 376.6KB