Delta Tips: How to Make a Print Layout (2)

  • Summary

NMDT_0069

Customize lists of acquisition and processing parameters to print

In Delta software, it is possible to create sophisticated layouts for printing and presentation.
This issue of Delta Tips explains additional functions of the Page Layout Editor. It is possible to customize various lists and text related to the sample and NMR data to print.
In the examples below, we will customize lists of acquisition parameters , processing parameters and integration data .

Customize lists

For the details on creating various layouts, refer to the Delta Tips issue titled How to Make a Print Layout (1) (⇒NMDT_0068 ).

① Select a data. Select a data box and click the Select Data button in the Page Layout Editor window.
And then select the data with the cursor to load this data to the data box. Note that the name of the selected data is displayed in the data box.

Page Layout Editor

If you wish to display lists of peaks and integrals, create the reports as follows:
Open a data and select Reports-Peak Report or Integral Report on the menu bar to create a peaks list or integrals list. Click the File Information button () in the Delta Console window to open the File Information window.
Note that two reports have been shown in the Reports tab below. One of them reports peaks, the other one integrals.

For the details on creating and saving peak lists, refer to the Delta Tips issue titled How to Create and Save Peak List (⇒NMDT_0014 ). For integral lists, refer to the Delta Tips issue titled How to Create and Save Integral List (⇒NMDT_0015 ).

creating and saving peak lists

② Create a data box to load the report. Select the data box and select the source data in the list to link. In the example below, we have linked Box 2 and Box 3 to the source data Box 1.

Create a data box

③ Select the data box and click the button to select Data/ Parameters/ a Report/ a Processing List. In the example below, we have set Data in Box 1, Report in Box 2 and Parameters in Box 3.

Select the data box

④ Select Import-Select Report to open the Select Report window. Select the report and click the Select button to load the report. In the example below, we have loaded an Integral List to Box 2.

Import-Select Report

It is also possible to display text (.txt) as follows:
Select Import-TEXT to open the Specify text file window.

display text

⑤ Click the Print button to print the data.

Print button

For the details on saving layout, refer to the Delta Tips issue titled How to Make a Print Layout (1) (⇒NMDT_0068 ).

For the details on setting features of data boxes, refer to the Delta Tips issue titled How to Make a Print Layout (3) (⇒NMDT_0070 ).

How to set font:

It is possible to set the size and style of letters in a data box as follows:
Select Box-Attributes to open the Attributes window.

For the details on the Attributes window, refer to the Delta Tips issue titled Page Layout Editor (Part 1) (⇒NMDT_0019 ).

How to set font

How to use templates:

It is possible to use preset templates as follows:
Select a template from the Template menu or click the Open Template button to open the Open File window.

Refer to the Delta Tips issue titled Page Layout Editor (Part 2) (⇒NMDT_0055 ) for more details

How to use templates
Please see the PDF file for the additional information.
Another window opens when you click.

PDF 1.28 MB

SEARCH APPLICATIONS

Related Products