- Observation of biological specimen by JEM-1400Flash ー Flow from sample preparation to observation ー
Observation of biological specimen by JEM-1400Flash ー Flow from sample preparation to observation ー
- Summary
EM2021-01
Introduction
Using an electron microscope allows for observations of micro-structures in a cell which is not possible with an optical microscope. In order to understand the detailed structure of organelles such as mitochondria and chloroplasts, an electron microscope is a powerful tool. However, with TEM, observation of a specimen in a living state is not possible as the observation is performed in vacuum. Also, in order for an electron beam to transmit, specimens need to be thinly sliced.
For this purpose, specimen preparation is a very important step to grasp the structure of the target without artifact. This application note describes the flow of sample preparation for biological specimens, using an example of a plant tissue to introduce data that is obtained by JEM-1400Flash.
Flow of specimen preparation
There are various methods in preparing biological specimens. As examples, we introduce specimen fixation by using a chemical fixation method and ultra-thin sectioning method by using an ultramicrotome. The specimen preparation process follows this order: 1. Specimen cutting, 2. Fixation, 3. Dehydration, 4. Substitution, 5. Embedding, 6. Polymerization, 7. Trimming, 8. Ultra-thin sectioning, 9. Staining. Each process is explained below.
1. Specimen cutting
The specimen is cut into small pieces using a razor blade, so that the chemicals to be used in further processing can easily permeate.
2. Fixation
A process to stop structural change of biological specimens with only a few artifacts. For chemical fixation, fixation is performed twice: prefixation and postfixation. Prefixation is used to fix proteins, and postfixation is used to fix lipids.
Note: Osmium tetroxide is highly volatile and reactive, and can damage the respiratory system, skin, and mucous membranes of the user. So, the process should be performed in a fume food.
Reagents
- Prefixation solution:2.5 % glutaraldehyde and 2 % paraformaldehyde in 0.1 M HEPES*¹ (half karnovsky)
- Postfixation solution:1 % osmium tetroxide solution (OsO4 in 0.1 M HEPES)
- Wash solution:0.1 M HEPES or distilled water
- Sample bottle
*1 HEPES: an abbreviation of buffer solution 2- [4- (2-Hydroxyethyl) -1-piperazinyl] ethanesulfonic acid
3. Dehydration
Use ethanol or acetone to dehydrate water content in the cells. If moisture remains, the remaining water content will prevent polymerization of resin during subsequent resin embedding. Follow the flow below to soak the specimen in each concentration of ethanol to substitute the water content with ethanol.

4. Substitution
When dehydrated with ethanol, it is substituted with propylene oxide (PO) or similar as an intermediate agent to blend with the epoxy resin.
Reagents
- PO
- Epoxy resin
※TAAB’s EPON 812*2 was used.
- PO only 2x5 min
- PO:resin=2:1 2 hours
- PO:resin=1:1 2 hours
- PO:resin=1:2 2 hours
- Resin only 18 to 24 hours
- Resin adjustment -
- Measure EPON 812*2, DDSA*3, MNA*4 in the proportions shown in the table and put them in a beaker.
- Stir well using a stirrer
- Add DMP-30*5 and stir
- Degas the resin and remove air bubbles
Table: Epoxy resin compounding
(%) | soft | medium | hard |
---|---|---|---|
EPON812 | 48 | 48 | 48 |
DDSA | 30 | 19 | 12 |
MNA | 20 | 33 | 40 |
DMP-30 | 2 | 2 | 2 |
Reference:TAAB’s DATA SHEET 12a
*2 EPON 812:resin, *3 DDSA・*4 MNA:hardener, *5 DMP-30:accelerating agent
5. Embedding
In order to prepare an ultra-thin section, it is necessary to embed the specimen in hard embedding agent such as resin.
Liquid resin will be cured during the polymerization process later.
Tools
- Epoxy resin
- Silicon embedding plate (Fig. 1)
- Pour the resin
- Place the specimen at the tip of the silicon embedding plate
- Let it sit for a while to blend

Fig. 1 Silicon embedding plate
6. Polymerization
Heat the resin to cure, to allow for thin slicing of the resin. Place the silicon embedding plate in an oven and keep it at 60°C for three days. Stabilizing the temperature during polymerization can prevent polymerization failure of the resin.
Tools
- Oven for resin polymerization
7. Trimming
Trim the specimen under the optical microscope so that the area to be observed has the size to be placed on the specimen grid and shape the surface to be observed.
Tools
- Razor blade
- Optical microscope

8. Ultra-thin sectioning
Remove the resin to expose the surface of tissue using a glass knife. Then cut into slices thin enough to transmit an electron beam with a diamond knife and an ultramicrotome, and placed on a grid.
Tools
- Ultramicrotome
- Grid
- Diamond knife
- Glass knife
- Eyelash probe(Fig. 2)

Fig. 2 Eyelash probe

9. Staining
This operation is used to enhance the contrast of biological specimens which contain many light elements. The scattering contrast is enhanced by binding heavy elements to the specimen. When staining the ultra-thin sections on a grid, double-staining is performed using uranyl acetate and lead citrate as staining agents. Uranyl acetate stains nucleoplasm and ribosome, while lead citrate stains cell membrane, glycogen granules, and ribosome.
Reagents
- Uranyl acetate*6
- Lead citrate
*6 Uranyl acetate is a compound which is an internationally controlled substance that can only be used in licensed facilities.
Alternative solutions to uranyl acetate include ytterbium acetate.
Result
Fig. 3 shows the results of the observation of an ultra-thin section specimen prepared by following the above flow, by using a TEM (JEM-1400Flash).
- Please see the PDF file for the additional information.