Close Btn

Select Your Regional site

Close

Analysis of polydimethylsiloxanes by GC/TOFMS [GC-TOFMS Application]

MS Tips No.131 The author would like to thank Mr.Junji Takaie of Showa Denko K.K. for his generous gift of the samples.

Three polydimethylsiloxane compounds; octamethylcyclotetrasiloxane (sample No. 1), decamethylcyclopentasiloxane (No. 2) and octa(dimethylsiloxy)silsesquioxane (No. 3), were analyzed using JMS-T100GC. The samples No. 1 and No. 2 were relatively small molecules with molecular weights of 296 and 370 respectively and could be introduced through the mass reference sample inlet (reservoir.) They can be used as internal mass references for accurate mass measurement. The sample No. 3 has a molecular weight of 1,016 and was difficult to be introduced through the reservoir. Sample No. 3 was analyzed by GC/MS.
Please see the PDF file for the additional information.
Another window opens when you click.

PDF 162KB

Related Products

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Science Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.