Electrostatic Dose Modulator (EDM)

  • Features
  • Specifications
  • Application
  • Related Products
  • Information

Features

Electrostatic Dose Modulator (EDM)
  • Lightning-fast speeds:

    EDM systems achieve switching times faster than 20 ns.
  • Independent intensity adjustment:

    By rapidly turning the beam on and off with variable pulse widths, the EDM makes it easy to adjust the average beam intensity without changing the image conditions. A desktop accessory knob provides an intuitive interface to adjust the dose attenuation factor.
  • Dose structuring:

    Users can take control of their illumination by applying dose in pulses with variable durations as short as 100 ns and frequencies up to 1 MHz.
  • Modern control software:

    The EDM works out of the box as a fast beam blanker. Control software is included for programmable dose attenuation, logging, and more.
  • Integration:

    The EDM act as a fast reliable pre-sample beam blanker while supporting a companion Relativity Sub-Framing System as well as third-party hardware.

Pulsed beam TEM illumination*

pulsed beam TEM illumination

Pulsed beam STEM Illumination*

Pulsed beam STEM illumination

Dose attenuation by pulsed illumination with Frequency of 500 kHz (2 μs) in a high resolution STEM at 300 kV using Si[110] in the condition of pixel dwell time with 19 μs/pixel (1,024 x 1,024 pixels), by changing duration ration from 90% to 50% during acquisition of one STEM image in 20 s.

* Electrostatic Dose Modulator (EDM) offers two versions, EDM Basic and EDM Advanced. EDM Basic provides Pulsed Beam TEM Illumination, and EDM Advanced does both Pulsed Beam TEM Illumination and Pulsed Beam STEM Illumination.

Specifications

Fast beam blanking

Quantity Value
Maximum pulse frequency 1 MHz
Transition time 90%-10% < 20 ns
Blanking signals 3 inputs

Dose attenuation

Configuration EDM Basic EDM Advanced
Application TEM imaging TEM and STEM, pump-probe dynamics
Pulse repetition frequency (max) 1 MHz
Pulse width increment 62.5 ns 10 ns
Minimum pluse width 125 ns 100 ns
Synchronization No sysnchronization TTL trigger input

Applicable model: JEM-ARM300F2, JEM-ARM200F (CFEG), NEOARM (CFEG), JEM-F200 (CFEG), JEM-2200FS*, JEM-2100F*
(* These models should be equipped with a standard column without the Cs corrector.)

Gallery

Related Products

Information

Installation Examples