• Features
  • Specifications
  • Application
  • Related Products
  • Information

Features

IB-09060CIS Cryo Ion Slicer™

For easy preparation of TEM thin film samples of heat sensitive materials

  • Minimizes thermal damage during ion irradiation using liquid nitrogen
  • Long cryo retention time to fully eliminate damage
  • Specimen exchange with liquid nitrogen injected

*Maximum sample size: 2.8 mm (L) x 1.0 mm (W) x 0.1 mm (T)

Catalogue Download

Specifications

IB-09060CIS
Ion accelerating voltage 1 to 8kV
Tilt angle Up to 6°(0.1°step)
Ion-beam diameter 500 µm or more (full width at half maximum)
Milling speed 5 µm/min(accelerating voltage: 8 kV, Si equivalent value)
Gas Argon
Maximum specimen size 2.8mm(L)×1.0 mm(W)×0.1mm(T)
試料冷却 Specimen-holdercooling temperature -120℃
Time for cooling attainment 1h
Cooling-retention time 8h
Specimen take-out time 30min
Pressure measurement Penning gauge
Main vacuum pump Turbo-molecular pump
CCD Camera Built-in
Dimensions and masses Basic unit 500mm(W)×600mm(D)×542mm(H)、70kg
Rotary pump 150mm(W)×427mm(D)×230.5mm(H)、16kg
Display 326mm(W)×173mm(D)×380mm(H)、3.7kg

Installation Requirements

IB-09060CIS
Power Single-phase AC 100 to 120 V±10 %, 50/60 Hz,0.5 to 0.6 kVA
Grounding 100Ω or less
Argon gas Pressure: 0.15±0.05 MPa (1.0 to 2.0 kg/cm2)Purity: 99.9999% or more (argon gas, gas cylinder and regulator should beprovided by the customer)
Hose joint: JIS B0203, Rc 1/4
Room temperature 20 to 25℃
Humidity 60% or less

* Specifications subject to change without prior notice.

Gallery

Related Products

Information

Installation Examples