• Features
  • Specifications
  • Application
  • Related Products
  • Information

Features

JEM-2200FS Field Emission Electron Microscope
The JEM-2200FS, a state-of-the art analytical electron microscope, is equipped with a 200kV field emission gun (FEG) and the in-column energy filter (Omega filter) that allows a zero-loss image, where inelastic electrons is eliminated, resulting in clear images with high contrast. And energy-filtered images forming with electrons at low loss or core loss energy provide chemical state or elemental information of a sample. Also, spectroscopy for elemental analysis and chemical analysis of specimens is available.
 

In-column energy filter (Omega filter)

The in-column energy filter enables you to obtain energy-filtered images and electron energy loss spectra. The optimally designed filter provides distortion-free filtered images.

Control system

The main components of the JEM-2200FS, such as the optical system, goniometer stage and evacuation system, are fully PC-controlled. This system stably produces high-quality data.

Imaging system

A new imaging system, consisting of four-stage intermediate lenses and two-stage projector lenses, achieves rotation-free energy-filtered TEM images and diffraction patterns over a wide range of magnifications and camera length.

Piezo-controlled goniometer

A new goniometer stage that incorporates a piezo device offers smooth operation for searching fields of view at the atomic level.

Integration to other instruments

The microscope can be fully controlled with PC.  The design concept enables us to integrate EDS system and CCD cameras.

Catalogue Download

Specifications

Configuration*1 Ultrahigh resolution
(UHR)
High resolution
(HR)
High specimen tilt
(HT)
Cryo
(CR)
High Contrast
(HC)
Resolution
 Point
 Lattice
0.19 nm
0.1 nm
0.23 nm
0.1 nm
0.25 nm
0.1 nm
0.27 nm
0.14 nm

0.31 nm
0.14 nm

Energy Resolution 0.8 eV(zero-loss FWHM)
Acc. Voltage 160 kV,200 kV*2
 Minimum step size
 Energy shift
50 V
3,000 V maximum (in 0.2 V steps)
Electron source
 Emitter ZrO/W(100) Schottky
 Brightness ≧4×108 A/cm・ sr
 Vacuum ×10-8 Pa order
 Probe current 0.5 nA at 1 nm probe
Power Stability
 Acc. voltage ≦1×10-6/min
 OL current ≦1×10-6/min
 Filter current ≦1×10-6/min
Objective Lens
 Focal length 1.9 mm 2.3 mm 2.7 mm 2.8 mm 3.9 mm
 Spherical
 aberration
0.5 mm 1.0 mm 1.4 mm 2.0 mm 3.3 mm
 Chromatic
 aberration
1.1 mm 1.4 mm 1.8 mm 2.1 mm 3.0 mm
 Minimum step 1.0 nm 1.4 nm 1.8 nm 2.0 nm 5.2 nm
Spot Size (diameter)
 TEM mode 2 to 5 nm 7 to30 nm
 EDS mode 0.5 to 2.4 nm 1.0 to2.4 nm - 4 to20 nm
 NBD mode - -
 CBD mode 1.0 to2.4 nm -
CB Diffraction
 Convergence
  angle(2α)
1.5 to 20 mrad or more -
 Acceptance angle ±10 ° -
Magnification
 MAG mode ×2,000 to 1,500,000 ×2,000 to 1,200,000 ×1,500 to1,000,000 ×1,200 to600,000
 LOW MAG mode ×50 to1,500
 SA MAG mode ×10,000 to800,000 ×8,000 to 600,000 ×8,000 to500,000 ×5,000 to 400,000
Field-of view size for energy selected image 80 mm dia. on final image plane (film) when 10 eV is selected
25 mm dia. on final image plane (film) when 2 eV is selected
Camera Length
 SA diffraction 150 to1,500mm 200 to2,000 mm 250 to2,500 mm 300 to3,000 mm
EELS Dispersion
 On slit 1.2 μm/eV at 200 kV
 On final image plane 50 to300 μm/eV at 200 kV
Specimen chamber
 Shift(XY / Z) 2mm/0.2mm 2mm/0.4mm 2mm/0.4mm 2mm/0.4mm 2mm/0.4mm
 Specimen tilt X / Y *3 ±25°/±25° ±35°/±30° ±42°/±30° ±15°/±10° ±38°/±30°
 Specimen tilt angle(X)*4 ±25° ±80° ±80° ±80° ±80°
EDS*5
 Solid angle*6 0.13 sr - 0.09 sr
 Take-off angle 25° - 20°
 

*1 : Select either configuration when ordering the JEM-2200FS.
*2 : 80 kV, 100 kV and 120 kV become possible when using the optional short-circuit switches for the accelerating tube.
*3 : When the Specimen Tilting Holder is used.
*4 : When the High Tilt Specimen Retainer is used.
*5 : An optional EDS detector is needed.
*6 : When 30 mm2 detector is installed.

Gallery

Related Products

Information

Installation Examples