JST-F series Electron Beam Power Supply
- Features
- Specifications
- Application
- Related Products
- Information
Features

The JST-F series offers excellent stability and has been the choice of many users.
Related product links
Specifications
Model | JST3F | JST-10F | JST-16F | |
---|---|---|---|---|
Output | Maximum output | 6.4kW | 10kW | 16kW |
Accelerating voltage | -4kV to -8kV | -4kV to -10kV | ||
Emission current | 0 to 0.8A | 0 to 1A | 0 to 1.6A | |
Filament current | Maximum 40A | |||
Accelerating voltage ripple rate | 5% or less (maximum output) | |||
Deflection coil output | 0 to 2A | |||
Sweeping coil output | Maximum frequency 500 Hz(※), Maximum output 8 Ap-p, ± 2 A DC | |||
Interlock | External: 3 points, Internal 5 points | |||
External signal | INPUT : 16 types, OUTPUT: 9 types | |||
Power supply input (50/60 Hz) | 3 phase 200 V ± 10%, 10 kVA | 3 phase 200 V ± 10%, 15 kVA | 3 phase 200 V ± 10%, 25 kVA | |
External dimensions | Power supply main unit | 570mm(W)×760mm(D)×1,545mm(H) | ||
Operation panel | 481mm(W)×300mm(D)×150mm(H) | |||
Mass | Approximately 360kg | Approximately 390kg | Approximately 450kg | |
Grounding | Ground resistance 10 Ω or less |
Attachments
ST-AS10F/ST-AS10G Arc Suppressor
Suppresses arcing during electron beam evaporation. Suppresses noise during arcing, and prevents interruption of the vapor deposition.ST-AS10F has a 2 stage arcing suppression circuit, and ST-AS10G has a 1 stage arcing suppression circuit.

ST-AS10G
Model | ST-AS10F | ST-AS10G |
---|---|---|
Applicable accelerating voltage/ current | -4kV 〜 -10kV 1.6A 以下 | |
Operating pressure | 6 × 10-2 Pa 以下 | |
Interruption time of the electron beam | 0.5ms or less* | 0.1ms or less |
Accelerating voltage overshoot during recovery | 20%/3 ms or less | |
Increased current of the rectifier wave during arcing | 1 A, 100 μS or less |
ST-203 Simultaneous Multi Electron Beam Source Unit
This is used when 2 electron beam sources will be used simultaneously with 1 power supply. There will be 2 operation panels, each one is assigned for dedicated operation of one of the 2 electron beam sources.
It is possible to use different types of electron beam sources,but there may be limitations on the accelerating voltage range, etc. depending on the combination.
ST-EXTSIG External Signal Control
Incorporated in to the operation panel. In addition to the standard 9 types of external control signals, enables the external output of various other signals, such as the emission current confirmation signal
Cable configuratio

Type | Name | Model | Length |
---|---|---|---|
W1 W2 W8 |
E1/E2/ grounding | BS-64410L06W128 | 6m |
ST-10MW128 | 10m | ||
ST-15MW128 | 15m | ||
W3 W4 |
Coil/HV (2 coiles) | BS-64420L06W34 | 6.5m / 5.5m |
ST-10MW34 | 10m | ||
ST15MW34 | 15m | ||
W5 | Emission return | BS-64430L06W5 | 5.5m |
ST-10MW5 | 10m | ||
ST-15MW5 | 15m | ||
W6 W7 |
Interlock/EB source ON | BS-64440L06W67 | 6m |
ST-10MW67 | 10m | ||
BS-64540L15W67 | 15m | ||
W9 | Remote | BS-64450L06W9 | 6m |
ST-10MW9 | 10m | ||
ST-15MW9 | 15m |