GC-QMS Application:Improving efficiency in drinking water analysis by high-end GC-QMS ~ Scan quantification of VOCs, 2-MIB&Geosmin, haloacetic acids, formaldehyde, Phenol by double column connection ~
- Summary
MSTips No.346
1.Introduction
Of the drinking water quality standards in Japan, the GC-MS method is adopted as the inspection method for 19 test items. The 19 test items covered by the GC-MS method are classified into 5 groups as volatile organic compounds (VOCs), substances causing musty odors (2-methylisoborneol, geosmin), haloacetic acids, formaldehyde, and phenols. One or more official methods have been established for each group. We already have reported an analysis application using the same column for the purpose of improving operational efficiency in these water quality analyses. The same column analysis of VOCs, 2-MIB, and geosmin using a mid-polar column was reported in MSTips No.334. The same column analysis of haloacetic acids, formaldehyde, and phenols using a non-polar column was reported in MSTips No.325.
JEOL Ltd. released the 6th generation high-end GC-QMS "JMS-Q1600GC UltraQuad™ SQ-Zeta" in 2021. We report on the construction of a system that enables measurement of all GC-MS items of water quality standards with a single instrument using the JMS-Q1600GC. The system can measure all test items without the need for column replacement by connecting a mid-polar column and a non-polar column to one JMS-Q1600GC in a double column configuration. In addition, by using the new attachment "High Performance Ion Source (EPIS)", an easier "Scan quantification" is possible instead of the "SIM quantification" that requires complicated condition settings. This MSTips reports the results of the evaluation of the linearity of the calibration curves for each item and the repeatability at a lower concentration limit.

Figure 1. Measurement items by GC-MS method in Japanese drinking water regulation
2. Measurement
2.1. Sample Preparation
VOCs: In a headspace vial containing 3 g NaCl and 10 mL of purified water, VOCs except 1,4-dioxane were added at 0.1, 0.2, 0.5, 1, 2, 5, and 10 μg / L and 1,4-dioxane at 1, 2, 5, 10, 20, 50, and 100 μg / L, respectively, for adjustment. For internal standards, fluorobenzene and p-bromofluorobenzene were added to each measurement sample at 2.5 ppb and 1,4-dioxane-d8 at a concentration of 200 μg / L. 2-Methylisoborneol & Geosmin: In a headspace vial containing 4.5 g NaCl and 10 mL of purified water, 2-methylisoborneol (2-MIB) and geosmin were added at 1, 2, 5, and 10 ng / L, respectively, for adjustment. The internal standard was 2,4,6-trichloroanisole-d3 added to a concentration of 20 ng / L. Haloacetic acids: For methyl chloroacetate, methyl dichloroacetate, and methyl trichloroacetate, the concentrations were adjusted by stepwise dilution with MTBE to 0.002, 0.004, 0.008, 0.02, and 0.04 mg / L as converted to the concentration of haloacetic acids in the test water. For the internal standard, 1,2,3-trichloropropane was added to each measurement sample to a concentration of 0.1 mg / L. Formaldehyde: PFBOA-formaldehyde was adjusted by stepwise dilution with n-hexane to 0.002, 0.001, 0.005, 0.01, 0.05, and 0.1 mg / L as converted to the concentration of formaldehyde in the test water. For the internal standard, 1-chlorodecane was added to each measurement sample to a concentration of 0.1 mg / L. Phenols: Phenol, 2-chlorophenol, 4-chlorophenol, 2,4-dichlorophenol, 2,6-dichlorophenol, 2,4,6-trichlorophenol were added stepwise with ethyl acetate to reach 0.0005, 0.001, 0.005, 0.01 mg / L as the concentration in the test water. After dilution, 50 μL of N,O-bis (trimethylsilyl) trifluoroacetamide was added to 1 mL of the aliquoted solution and allowed to stand for 1 hour. As an internal standard, acenaphthene-d10 was added to a concentration of 0.2 mg / L.
2.2. Measurement conditions
The measurement conditions for the samples are shown in Table 1. Scan mode was used for the MS data acquisition. Although Scan mode is less sensitive than SIM mode, it simplifies the workflow in analytical operations because it does not require the creation of measurement conditions necessary for SIM. The use of the EPIS in this MSTips enables both high-sensitivity measurement and operational cost reduction.
Table 1. Measurement condition
VOCs | Geosmin, 2-MIB | haloacetic acids | Formaldehyde | Phenol | ||
---|---|---|---|---|---|---|
HS | Sample temp. | 70°C | 80°C | |||
Heating time | 30 min | |||||
Sampling mode | Trap (3 times) | |||||
GC | Column | DB-1301 (Agilent Technologies, Inc.), 60 m × 0.32 mm id, 1 μm film thickness |
InertCap 1MS (GL Sciences Inc. ), 30 m × 0.25 mm id, 1 μm film thickness |
|||
Oven | 40°C for 3 min, to 100°C at 5°C / min, to 250°C at 10°C / min, and hold for 5 min |
40°C for 8 min, to 250°C at 15°C / min, and hold for 3 min |
50°C for 1 min, to 250°C at 15°C / min, and hold for 5 min |
70°C for 1 min, to 250°C at 15°C / min, and hold for 5 min |
||
Carrier gas | 83.44 kPa (Constant Pressure) | 1 mL / min (Constant Flow) | ||||
Inlet temp. | 250°C | |||||
Injection mode | Splitless | |||||
Injection volume | 2 μL | |||||
MS | Interface temp. | 250°C | ||||
Ion source temp. | 250°C | |||||
Ionization current | 50 μA | 100 μA | 100 μA | 50 μA | 100 μA | |
Ionization energy | 70 eV | |||||
Acquisition mode | Scan | |||||
Scan range | m/z 45 ~ 200 | m/z 80 ~ 230 | m/z 40 ~ 160 | m/z 33 ~ 230 | m/z 33 ~ 300 |
3. Result
Table 2 shows the correlation coefficient of the calibration curve and the coefficient of variation (= C.V.) of the quantification value when the lower concentration limit of the calibration curve was continuously measured at n=5 for the measured component. Figure 2 shows the calibration curve for 1,4-dioxane (VOCs), 2-methylisoborneol, chloroacetic acid (haloacetic acids), and 2,4,6-trichlorophenol (phenols), and Figure 3 shows the chromatogram of the lower concentration limit of the calibration curve. The correlation coefficients shown in Table 2 are greater than 0.999 for all components, and good linearity is obtained in the concentration range adjusted in this MSTips. The coefficients of variation at the lower concentration limit were also less than 5% for all components, indicating that it is possible to sufficiently measure concentrations of 1/10 of the standard value, which is the sensitivity required for water quality testing.
Table 2. Correlation Coefficient and Coefficient of Variation (C.V.) of each compound.
Compound Name | Correlation Coefficient |
C.V. (%) |
Sample Conc. (μg/L) |
Standard Value (μg/L) |
---|---|---|---|---|
Carbon tetrachloride | 0.9998 | 2.0 | 0.1 | 2 |
1,4-dioxane | 0.9999 | 3.4 | 1 | 50 |
trans-1,2-dichloroethylene | 0.9999 | |||
cis-1,2-dichloroethylene | 0.9999 | |||
1,2-dichloroethylene | 0.8 | 0.2 | 40 | |
Dichloromethane | 0.9997 | 3.4 | 0.1 | 20 |
Tetrachlorethylene | 0.9998 | 0.9 | 0.1 | 10 |
Trichlorethylene | 0.9999 | 1.4 | 0.1 | 10 |
benzene | 0.9999 | 1.0 | 0.1 | 10 |
Chloroacetic acid | 0.9998 | 2.6 | 2 | 20 |
Chloroform | 0.9999 | 1.0 | 0.1 | 60 |
Dichloroacetic acid | 0.9999 | 1.9 | 2 | 30 |
Dibromochloromethane | 0.9997 | 2.6 | 0.1 | 100 |
Compound Name | Correlation Coefficient |
C.V. (%) |
Sample Conc. (μg/L) |
Standard Value (μg/L) |
---|---|---|---|---|
Total trihalomethane | 1.6 | 0.4 | 100 | |
Trichloroacetic acid | 0.9997 | 0.8 | 2 | 30 |
Bromodichloromethane | 0.9999 | 1.8 | 0.1 | 30 |
Bromoform | 0.9990 | 2.5 | 0.1 | 90 |
Formaldehyde | 0.9999 | 0.7 | 1 | 80 |
2-Methylisoborneol | 0.9992 | 4.3 | 0.001 | 0.01 |
Geosmin | 0.9995 | 1.6 | 0.001 | 0.01 |
Phenol | 0.9988 | 0.9 | 0.5 | 5 |
2-Chlorophenol | 0.9998 | 0.5 | 0.5 | |
4-Chlorophenol | 0.9998 | 0.6 | 0.5 | |
2,6-dichlorophenol | 0.9999 | 0.9 | 0.5 | |
2,4-dichlorophenol | 0.9999 | 0.2 | 0.5 | |
2,4,6-Trichlorophenol | 0.9999 | 3.7 | 0.5 |

Figure 2. Calibration curve of 1,4-Dioxane, 2-MIB, Chloroacetic acid, 2,4,6-Trichlorophenol.

Figure 3. SIM chromatogram of 1,4-Dioxane, 2-Methylisoborneol, Chloroacetic acid, 2,4,6-Trichlorophenol at minimum plot of each calibration curve.
4. Conclusion
The JMS-Q1600GC UltraQuad™ SQ-Zeta, equipped with an EPIS, is connected to a mid-polar column for VOCs, 2-MIB, and geosmin and a non-polar column for haloacetic acids, formaldehyde, and phenols in a double column configuration to create a GC-MS system that can measure all test items subject to GC-MS in water quality standards. This system enables the measurement of all GC-MS target items without the need for column replacement involving a vacuum shutdown, and the use of scan quantitation, which does not require complicated settings of measurement conditions, significantly reduces operation costs.
- Please see the PDF file for the additional information.